EasyManua.ls Logo

iDig 2D Touch - Page 29

iDig 2D Touch
187 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Procedures Bucket Calibration
Automatic Bucket Calibration - Dog Bone
Once the Sensor is in position press
“Next”
The system needs to detect which side of
the bucket the Sensor is installed. Raise
the dipper stick high enough off the
ground so that you can open the bucket.
Press “Next
Slowly close the bucket until the pop-up
screen appears. If the correct side is
determined, Press “OK” to confirm and
continue. Otherwise press “Cancel” and
repeat the process.
29