EasyManuals Logo

Industrial Data Systems IDS 430 User Manual

Default Icon
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #97 background imageLoading...
Page #97 background image
IDS440 Users Manual version 4.2 Set Operation Parameters
26
Parameter 4. Motion Detection Delay (6)
Motion detection delay is used by the motion detection function to determine if the scale
is in motion (not stable). Enter the amount of time (in tenths of a second) that the scale
must be stable for motion status to be false (scale status = stable).
Example: 0 = disables motion detection.
2 = 0.2 second delay before motion = false (stable = true).
15 = 1.5 second delay.
30 = 3.0 second delay.
Parameter 5. Motion Detection Band (6)
The motion band is used by the MOTION DETECT function to determine if the scale is
in motion (not stable). If the amount of change between display updates (parameter 2) is
greater than the motion band, a MOTION status is generated (unstable condition). The
motion band is entered in units of ¼ dd, i.e.. For a motion band of:
½ display division enter 2 ½ x 4 = 2.
2 display divisions enter 8 2 x 4 = 8.
4.5 display divisions enter 18 4.5 x 4 = 18.
Parameter 6. Display Blanking During Motion (No Blanking)
When display blanking is enabled, the display is blanked (turned off) when the scale is in
motion (not stable). Enter a 1 (yes) to enable display blanking, enter a 0 (no) to disable
display blanking.
Parameter 7. Zero Tracking Delay (0)
The automatic zero tracking function removes small deviations from zero. The IDS440
adjusts the zero setting by approximately 1/30
th
of a display digit per delay interval. The
delay interval is the number effective A/D conversions between adjustments.
Parameter 8. Zero tracking band (12)
The zero tracking band is the maximum displacement from zero wherein zero tracking is
active. In order for zero tracking to be active, the following conditions must be met:
Scale weight is within Zero Track Band.
Scale is stable (no motion)
Zero Track Band is greater than 0 (zero).
The zero tracking bands are entered in units of ¼ dd, i.e.. For a band of:
½ display division enter 2 ½ x 4 = 2.
2 display divisions enter 8 2 x 4 = 8.
4.5 display divisions enter 18 4.5 x 4 = 18
Parameter 9. Push Button Zero Percent (100)
Enter the percent of full-scale capacity that can be acquired as the zero point with the
ZERO key. 1 = 1%, 2 = 2%, ... 100 = 100%.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Industrial Data Systems IDS 430 and is the answer not in the manual?

Industrial Data Systems IDS 430 Specifications

General IconGeneral
BrandIndustrial Data Systems
ModelIDS 430
CategoryMeasuring Instruments
LanguageEnglish