IDS 430/433 Users Manual Version 1.0
- 32 -
Parameter 4. Motion Detection Delay (6)
Motion detection delay is used by the motion detection function to determine if
the scale is in motion (not stable). Enter the amount of time (in tenths of a
second) that the scale must be stable for motion status to be false (scale status =
stable).
Example: 0 = disables motion detection.
2 = 0.2 second delay before motion = false (stable = true).
15 = 1.5 second delay.
30 = 3.0 second delay.
Parameter 5. Motion Detection Band (6)
The motion band is used by the MOTION DETECT function to determine if the
scale is in motion (not stable). If the amount of change between display updates
(parameter 2) is greater than the motion band, a MOTION status is generated
(unstable condition). The motion band is entered in units of ¼ dd, i.e.. For a
motion band of:
½ display division enter 2 ½ x 4 = 2.
2 display divisions enter 8 2 x 4 = 8.
4.5 display divisions enter 18 4.5 x 4 = 18.
Parameter 6. Display Blanking During Motion (No Blanking)
When display blanking is enabled, the display is blanked (turned off) when the
scale is in motion (not stable). Enter a 1 (yes) to enable display blanking, enter a
0 (no) to disable display blanking.
Parameter 7. Zero Tracking Delay (0)
The automatic zero tracking function removes small deviations from zero. The
IDS 430/433 adjusts the zero setting by approximately 1/30
th
of a display digit per
delay interval. The delay interval is the number effective A/D conversions
between adjustments.
Parameter 8. Zero tracking band (12)
The zero tracking band is the maximum displacement from zero wherein zero
tracking is active. In order for zero tracking to be active, the following conditions
must be met:
Scale weight is within Zero Track Band.
Scale is stable (no motion)
Zero Track Band is greater than 0 (zero).
The zero tracking bands are entered in units of ¼ dd, i.e.. For a band of:
½ display division enter 2 ½ x 4 = 2.
2 display divisions enter 8 2 x 4 = 8.
4.5 display divisions enter 18 4.5 x 4 = 18

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Industrial Data Systems IDS 430 and is the answer not in the manual?

Industrial Data Systems IDS 430 Specifications

General IconGeneral
BrandIndustrial Data Systems
ModelIDS 430
CategoryMeasuring Instruments
LanguageEnglish