IDS 430/433
Weight Indicator
Installation/Calibration/Operation
Version 1.a
02/24/2002
Industrial Data Systems, Inc.
590 W. Freedom Ave., Orange, CA 92865-2530
Tech Support 714-921-1353
Website: www.industrialdata.com Email: Sales@industrialdata.com

Table of Contents

Questions and Answers

  • S
    Sandra PetersAug 11, 2025
    Why is my Industrial Data Systems Measuring Instruments printer not printing (serial RS232)?
    • S
      Stacy JordanAug 11, 2025
      If the Industrial Data Systems Measuring Instruments printer is not printing via serial RS232, there might be an issue with the cable connection or pin configuration. Check the cable connections and pin configuration. Also, the baud rate and format on serial port 1 might not match the printer. Adjust the baud rate and format on serial port 1 to match the printer. Finally, the transmitter/receiver IC (U3 MC145406) could be damaged, in which case, replace the damaged transmitter/receiver IC.
  • A
    Austin WestAug 17, 2025
    What to do if continuous output does not update on Industrial Data Systems Measuring Instruments?
    • K
      Kristin AlexanderAug 17, 2025
      To resolve a non-updating continuous output on your Industrial Data Systems Measuring Instruments, first ensure that 'TX CONTINUOUS' is enabled in configuration Parameter 28. Also, check for any cable connection problems or invalid pin configurations. The baud rate and format on serial port 2 must match the receiver; adjust if necessary. Lastly, a damaged transmitter/receiver IC (U3 MC145406) could be the cause, requiring replacement.
  • E
    Emma MasonAug 23, 2025
    How to troubleshoot input port 2 not working on Industrial Data Systems Measuring Instruments?
    • S
      Steve PerezAug 23, 2025
      To fix a non-working input port 2 on your Industrial Data Systems Measuring Instruments, start by ensuring the baud rate and format on serial port 2 match the host. Check the cable connections and pin configuration for any issues. Verify that the internal jumper JP3 (RS232/422/485) is properly set. If problems persist, the RS232 transmitter/receiver IC (U3 MC145406) or the RS422/485 transmitter/receiver IC (RS422 {U11 75176} RS485 {U7 75176}) might be damaged and need replacement.
  • Y
    yvonnecrossAug 29, 2025
    What to do if input port 1 is not working on Industrial Data Systems Measuring Instruments?
    • J
      John HendricksAug 29, 2025
      To address a non-working input port 1 on your Industrial Data Systems Measuring Instruments, ensure that the baud rate and format on serial port 1 matches the host. Also, verify that the cable connections and pin configuration are correct. If the issue persists, the RS232 transmitter/receiver IC (U3 MC145406) might be damaged and require replacement.
  • M
    Michael BlackSep 4, 2025
    Why is the weight display invalid or fluctuating on Industrial Data Systems Measuring Instruments?
    • S
      sromeroSep 5, 2025
      If the weight display is invalid or fluctuating on your Industrial Data Systems Measuring Instruments, it could be due to invalid calibration parameters; check and correct these parameters. Alternatively, there may be problems with the load cell or load cell input cable; inspect and resolve any issues.