7
Only Applicators manufactured or approved by InMode Ltd. should be used
with EvolveX System.
Connect the power cord to a properly polarized and grounded power source
with the frequency and voltage characteristics that match those listed on the
back of the unit.
Connect the System power cord to a properly grounded receptacle. Do not use
power plug adapters.
Always turn off and unplug the device before cleaning.
The patient should not come into contact with metal parts which are earthed
or which have an appreciable capacitance to earth. The use of antistatic sheeting
is recommended for this purpose. Treatment bed or chair should not be electric.
Use the lowest output setting necessary to achieve the desired treatment
effect. The higher RF energy is applied, the greater the possibility of unintended
thermal damage.
Failure of the equipment could result in an unintended increase of output
power.
The cables of the Applicator should be positioned in such a way that contact
with the patient or other leads is avoided.
Fire / Explosion Hazard - The following substances will contribute to increased
fire and explosion hazards in the operating room:
– Flammable substances (such as alcohol-based skin prepping agents and
tinctures).
– Naturally-occurring flammable gases which may accumulate in body
cavities such as the bowel.
– Oxygen enriched atmospheres.
– Oxidizing agents (such as nitrous oxide [N2O] atmospheres).
– Endogenous gases.
The energy and heating associated with the System can provide an ignition
source. Observe fire precautions at all times. When using EvolveX in the same
room with any of these substances or gases, prevent their accumulation or
pooling within the area where EvolveX procedures are performed.