Section 2: Safety
5
There are no user-serviceable parts in the System, and all service and repair must be
performed only by the factory or authorized field service technicians.
2.3 Cautions
The following cautions should be heeded for safe System use:
Do not touch the System’s inner parts.
Service is supplied by company authorized personal only.
To avoid damage, do not allow the Applicator to come in contact with hard
materials.
2.4 Electrical and Mechanical Safety
Keep all covers and panels of the system closed. Removing the covers creates a
safety hazard.
Keep hands away from the Applicator during the System start-up.
Perform maintenance procedures when the system is shut down and
disconnected from the power.
The system is grounded through the grounding conductor in the power cable.
This protective grounding is essential for safe operation.
Move the system slowly and carefully. The System weighs approximately 33kg
(72lb.) and may cause injury if proper care is not used when moving it.
Provide as much distance as possible between the system, RF Applicator and
other electronic equipment as the activated RF generator may cause interference
between them.
2.5 Fire Hazards
Materials conducting RF energy cause a temperature rise of the absorbing
material. Do not use the System in the presence of explosive or flammable
materials conductive to RF.
Do not use the system in the presence of explosive or flammable materials.
Materials conducting RF energy cause an increase in temperature of the
absorbing material.
Do not use flammable substances when preparing the skin for treatment. Be
especially careful with the use of oxygen.