▪ If alcohol is used for cleaning and disinfecting, it must be allowed to dry
thoroughly before the System is used.
2.6 Safety Features of the System
The system incorporates the following safety features. All personnel operating the
system should be familiar with these features.
▪ System has unique password to avoid device operation by non-authorized
personnel.
▪ An audible tone indicates energy activation.
▪ Special audible tone will be activated in case of bad coupling of one of the
applicators.
▪ Patient call button allows to pause the treatment in case of discomfort
▪ During activation, the system performs a self-test of the hardware.
▪ Hardware is tested every few msec. to ensure proper operation of electrical
circuits.
▪ Skin surface is monitored during the RF treatment. RF energy delivery is
terminated when skin temperature accidentally reaches the Cut-Off level.
▪ System starts at a low setting.
2.7 Safe Use of the Active Accessories
▪ Examine the connection of the Applicator through the connector to the system
before using. Improper connection may result in arcs and sparks, accessory
malfunction, or unintended treatment effects.
▪ Do not wrap the Applicator cords around metal objects. It may induce current
that could lead to electrical shocks, fire or injury to the patient or personnel.
▪ When using the RF applicators, ensure that all electrodes are in full contact with
the skin. Bad coupling of electrodes with the skin results in a specific warning
sound, a message on the screen, and disabling RF.
Do not connect a wet accessory to the System.
Do not immerse the Applicator under water at any time.
This equipment is for use only by qualified medical professionals trained in the
particular technique to be performed.