8.6 Treatment Recommendations
The caregiver shouldn’t leave the patient unattended during
treatment
The patient can use the Patient call button to pause the treatment if
the discomfort becomes excessive
The skin should be observed for excessive erythema or edema. If
observed, stop the treatment and move to the next treatment
zone.
1. Follow Device Start-Up from Section 5.
2. Set treatment parameters. Suggested starting treatment parameters
for Tite Applicator are shown in the table below:
3. Any combination of treatment parameters should be according to
skin response and patient tolerance.
4. Always start with low settings and observe the skin's reaction and
patient comfort before increasing the RF energy.
5. Apply conductive gel to the skin, addition gel can be applied directly
to the Tite units.
Prior to using water-based gel, ensure that the container of
ultrasound gel has not passed the expiration date.
6. Set Tite units on the belt.
7. Apply the belt with the Tite units on treatment area while the patient
is positioned in the same position as the treatment will be
performed.
8. Ensure there is minimal movement after the Tite units are applied.
9. There should be a complete contact between the unit and the skin
during the treatment session.