These indicators show online status for each of the connected
Applicators. The grey color indicates that Applicator is not
connected or not recognized.
4.4 Sound Indicator
▪ Periodic beeping signal is emitted, when energy is delivered.
▪ A warning sound tone indicates Bad Coupling.
4.5 Cut-Off Temperature Control
The cut-off temperature up to an upper limit of temperature of 43°C is constantly
maintained. When the measured temperature approaches the Cut-off Temperature,
the tone beeps double in speed. It becomes faster when the Cut-off Temperature is
reached and RF is instantly inactivated. As soon as the temperature drops below the
Cut-off temperature, RF starts again, thus maintaining the desired temperature with
safety and when reached, RF delivery is automatically stopped. Temperature is
monitored by a temperature sensor in the Applicator and serves as a safety feature.