Section 4: Device Description
22
applicators 1&4 and 2&3. The System starts up at minimal
intensity level setting.
Treatment Time
This indicator shows time when energy is applied sequentially to
the Applicators.
Applicators
Status Indicators
(1-4)
These indicators show online status for each of the connected
Applicators. The grey color indicates that Applicator is not
connected or not recognized.
Main Menu
Return to the Main Menu to select the Applicator.
4.4 Sound Indicator
Periodic beeping signal is emitted, when energy is delivered.
A warning sound tone indicates Bad Coupling.
4.5 Cut-Off Temperature Control
The cut-off temperature up to an upper limit of temperature of 43°C is constantly
maintained. When the measured temperature approaches the Cut-off Temperature,
the tone beeps double in speed. It becomes faster when the Cut-off Temperature is
reached and RF is instantly inactivated. As soon as the temperature drops below the
Cut-off temperature, RF starts again, thus maintaining the desired temperature with
safety and when reached, RF delivery is automatically stopped. Temperature is
monitored by a temperature sensor in the Applicator and serves as a safety feature.