Section 7: Tone Treatment Information
39
The following processes validated for the Tone Applicators when used in
accordance with the instructions provided for cleaning products and/or
processes.
Any deviation from said instructions or the cleaning agents listed below
may affect the performance or durability of the product, and it’s prohibited.
Review those instructions for additional warnings and cautions.
We recommend the usage of the following cleaning agent that was tested
in our company for material compatibility as 70% alcohol solution.
Cleaning Procedure
Clean the Applicator units thoroughly with alcohol absorbed pad (not
soaked) and repeat as necessary. Do not submerge in fluids. Leave it for
complete drying.
Pre-use Check
Before each use of the Applicator, the device must pass the following:
Check to ensure proper cleaning and drying of the Applicator.
Visually inspect the applicators for residual soil. Repeat the Cleaning
Procedure in the event that residual soil is visible.
Inspect all components of the Applicator for visible damage.
Do not use the applicator/s in case of unacceptable deterioration
was observed such as corrosion, discoloration, pitting, or cracked
seals.
7.6 Pre-Treatment Recommendations
During the patient’s first visit the treating attendant should:
Review all indications
Review all contraindications and precautions
Complete the medical history and physical prior to treatment
Sign the informed consent prior to the procedure.
Perform a test immediately prior to the procedure to visualize the area
that is going to be treated
Long and dense hairs may affect the treatment and may be
shaved according to physician’s discretion.