Evolve System Operator Manual
Section 11: System Specifications
53
▪ Use of the System with any accessory, transducer or cable other than
those specified may result in increased EMISSIONS or decreased
IMMUNITY than those specified.
▪ The System should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the System should
be observed to verify normal operation in the configuration in which it
will be used.
IEC 60601-1-2 Edition 4.0 (2014)
Environment of intended uses:
Professional Healthcare Facility Environment
Summary of Test Results
Emission (IEC 60601-1-2 section 7.1 & 7.2 & IEC 60601-2-2)
Radiated emission
Freq. range: 30 - 1000 MHz
Conducted emission
Freq. range:150 kHz - 30 MHz
Group 1 Class A
100/ 120/ 230 VAC
Harmonic current emission
test
Voltage changes, Voltage
fluctuations and Flicker test
Immunity (IEC 60601-1-2 section 8.9 & 8.10 IEC 60601-2-2)
Immunity from Electrostatic
discharge (ESD)
8 kV contact discharges &
15 kV air discharges
Immunity from radiated
electromagnetic fields
3.0 V/m
80 MHz 2.7 GHz, 80% AM, 1kHz
Immunity from
Proximity field from wireless
communications equipment
List of frequencies,
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
Immunity from Electrical Fast
transient (EFT)
± 2 kV on 230 VAC mains,
Pedal, Applicator
Tr/Th – 5/50 ns, 100 kHz
±1.0 kV DM/ 2.0kV CM on 230 VAC
Tr/Th – 1.2/50 (8/20) s
Immunity from conducted
disturbances induced by RF
fields
3.0 & 6.0 VRMS on 230 VAC mains,
Pedal, Applicator
0.15÷ 80 MHz, 80% AM, 1 kHz
Immunity from Voltage dips,
short interruptions and
voltage variations
230 & 100 VAC mains;
0 % - 10 ms; 70% - 500 ms;
0% - 20 ms; 0% - 5sec