Chapter: Maintenance
102 M10-17547-EN
If your testing application generates debris, particularly conductive or abrasive dust,
refer to “Ingress Protection” on page 34 for additional maintenance advice.
Load frame maintenance
Refer to the documentation provided with your original system for maintenance
procedures including daily checks, periodic inspections and testing limit stops.
Retrofit Controller maintenance
Replace a Fuse
Warnings
Electrical Hazard - Shut down the system and disconnect the power cable from the
main power source before replacing a fuse. Dangerous voltages can cause
personal injury.
Hazard - Do not remove covers to any component of your system, unless it is
specified in a procedure.
There are dangerous voltages and rotating machinery inside the machine that may
cause bodily injury or damage to equipment.
Caution
Replace a fuse with the same type and size as the original. Installing the wrong fuse
could damage the electrical circuits inside the machine.
1. Ensure that the power switch is in the Off (0) position and disconnect the power cable
from the power source. Verify that no LEDs are illuminated on the indicator panel.
2. Insert a small flat-head screwdriver into the middle of the connector and pry out the
fuse holder, as shown below.