15
68TM-Retrofit Power Requirements and Cords
Product Support: www.instron.com
68TM-Retrofit Power Requirements and Cords
Power requirements
For 230 Vac operation, the frame is normally set to 240 V.
Electrical hazard - to ensure safety and Electromagnetic Compatibility (EMC) the
facility’s receptacle must be a 3-wire grounded receptacle. The ground must be a
low impedance earth ground in accordance with national and/or local regulations.
You must use RCD (residual-current device), also known as GFCI (Ground Fault
Circuit Interruption), protection on all power sources if your testing system includes
water or any other fluid.
If your system includes any fluids (e.g. water-cooled grips, BioPuls baths, food testing
fixtures), you must use RCD protection on all power sources to protect the operator from
fluids that might leak into the load frame electronics.
Power cord selections
If you do not make a choice, Instron
®
chooses the customary power cord that matches
the shipping destination for the system.
Table 1. 68TM-Retrofit power requirements
Parameter Specification
Maximum Power (VA) 1400
Single Phase Voltage (Vac) (±10%) 100, 120, 220, 240
Frequency - Hz 47 to 63
Length of Power Cable - m (ft) 2.44 (8)