Disassembly of driving unit
Bring seat system in service position and arrest.
Remove both motor plugs at main module and also plug connection at voltage supply.
Remove cable binder at motor cable if necessary.
Prop up wheelchair that it is accessible from below and supply driving unit in addition to that.
Disassemble front wheels. See chapter 8.2.
Disassemble wheel hubs with brake discs. See chapter 8.3.
Release three screws (9) and remove mounting of brake caliper (8).
If necessary remove speedometer sensor carefully.
Remove screwing at bottom of shock absorber (7)so far, that the shock absorbers are freed from front axle
Remove screwing from thread rod (6) so far, that the thread rod is freed from gear.
Release screws (11) and nuts and remove rubber support (12) with sockets (13) and discs (14),
so that the driving unit is freed from gear rocker arms (10).
Remove driving unit downwards out of wheelchair.
Assembly
Assembly is carried out in reverse order.
CAUTION:
Note the weight of driving unit.
> Danger uf crushing<
NOTE:
The driving unit can only be completely replaced.
Repairwork at driving unit must be done by
INVACARE-service.
Setting in of carbon brushes
Set in carbon brushes carefully into the carbon brush mounting of motor.
Press feather of carbon brush with soldering cap carefully down, until it sticks in mounting.
Screw on closure cap.
NOTE:
The carbon bruh must not cant.