EasyManua.ls Logo

Invacare IRC5P - Page 14

Invacare IRC5P
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECTION 2—INSTALLATION/SEQUENCE OF OPERATION
Perfecto
2
14 Part No 1148070
Afterfiveminutes,iftheoxygenpurityexceeds85%±2%,theGREENlightwillcontinue
toilluminate.
Iftheoxygenlevelisnotabove85%±2%afterthefirstfiveminutes,thesystemwill
continuetomonitortheO
2
andwaitforamaximumof30minutesfromstartuptoreach
85%±2%beforeactivatinganalarm.Environmentalfactorssuchaslowvoltage,high
altitude,orageofthemachinewillaffectthetimerequiredtoreach85%±2%.
Iftheoxygenlevelisnotabove85%±2%withinthefirst30minutes,theoxygen
concentrationalarmsequencewillactivateandtheunitwillshutdown.
Whenoxygenconcentrationisabove85%±2%,thesensormeasuresoxygenpurityevery
10minutes.Ifareadingfallsbelow85%±2%,aYELLOWlightwillilluminate.Ifthe
oxygenpurityfallsbelow73%±3%theREDlight/Alarm/ShutDownmodewillactivate.

Table of Contents

Other manuals for Invacare IRC5P

Related product manuals