EasyManua.ls Logo

Invacare IRC5P - Page 30

Invacare IRC5P
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECTION 6—PREVENTIVE MAINTENANCE
Perfecto
2
30 Part No 1148070
ON EACH INSPECTION
MODEL NO. I RC SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY 4,380 HOURS
Check Oxygen Concentration
(UNITS WITHOUT SensO
2
)
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
(
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End of Compressor
*NOTE: Refer to Preventive Maintenance section of Service Manual.
NOTE:
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Check Oxygen Concentration
UNITS WITH SensO
2
)
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST
FIGURE 6.7 Preventative Maintenance Record

Table of Contents

Other manuals for Invacare IRC5P

Related product manuals