SECTION 11—REGULATOR
Perfecto
2
50 Part No 1148070
9. Adjustuntilpressurereadsasteadyfivep.s.i.±0.5p.s.i.
10. Oncerequiredpressureisachieved,reinstallcabinet.RefertoRemovingCabineton
page 23.
11. RefertoTroubleshootingonpage 16ifyouareunabletoadjustormaintainfivep.s.i.
±0.5p.s.i.
NOTE:Aftertenminutesofruntime,retestoutputpressureofconcentratortoinsureproper
operationand/orspeclevels.