TYPICAL PRODUCT PARAMETERS
Part No 1148070 9 Perfecto
2
*NOTE:Statedconcentrationlevelsachievedafterinitialwarmupperiod(approximately30
minutes).
IRC5PO2 INDICATORS
IRC5P INDICATORS
Electrical: No extension cords.
Placement: No closer than three inches from any wall, furniture, draperies,
or similar surfaces.
Tubing: 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO
NOT pinch).
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Recommended Storage and Shipping Temper-
ature:
-20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity.
LABEL
SYMBOL
O
2
PURITY INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O
2
over 85% (± 2%)
GREEN Indicator Light
O
2
Between 73% (± 3%)
to 85% (±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73% (±3%)
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down.
LABEL
SYMBOL
STATUS INDICATOR LIGHTS
(LED)
SYSTEM OKAY GREEN Indicator Light
SYSTEM FAILURE RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician
O
2
I/O

Table of Contents

Other manuals for Invacare IRC5P

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare IRC5P and is the answer not in the manual?

Invacare IRC5P Specifications

General IconGeneral
BrandInvacare
ModelIRC5P
CategoryMedical Equipment
LanguageEnglish

Related product manuals