SECTION 9—HANDLING AND REPLACING BATTERIES
TDX™ Wheelchairs 98 Part No 1114809
SECTION 9— HANDLING AND
REPLACING BATTERIES
WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may occur.
Warnings for Handling and Replacing Batteries
WARNING
Make sure power to the wheelchair is OFF before performing these procedures.
The use of rubber gloves is recommended when working with batteries.
Invacare strongly recommends that battery installation and battery replacement
always be done by a qualified technician.
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may result.
22NF batteries weigh 37 pounds each. GP24 batteries weigh 51 pounds each. Use
proper lifting techniques (lift with your legs) to avoid injury.
FORMULA TRE SEATING SYSTEMS, 2G TARSYS SYSTEMS, FORMULA PTO
SEATING SYSTEMS AND TDX WHEELCHAIRS WITHOUT POWERED
SEATING SYSTEMS ONLY - Use MK p/n M24SLDG or p/n M22NFSLDG batteries
only. Failure to use the correct battery size and/or voltage may cause damage to
your wheelchair and give you unsatisfactory performance.
FORMULA INVISIBLE SUPER LOW TILT SYSTEMS ONLY - Use MK p/n
M22NFSLDG batteries only. Failure to use the correct battery size and/or voltage
may cause damage to your wheelchair and give you unsatisfactory performance
ALWAYS use a battery lifting strap when lifting a battery. It is the most convenient
method and assures that the battery acid will not spill. It also helps to prolong the
life of the battery.
DO NOT tip the batteries. Keep the batteries in an upright position.
NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery
post(s) at the same time. An electrical short may occur and serious personal injury
or damage may occur.
When tightening the clamps, always use a box wrench. Pliers will “round off” the
nuts. NEVER wiggle the battery terminal(s)/post(s) when tightening. The battery
may become damaged.
The POSITIVE (+) RED battery cable MUST connect to the POSITIVE (+) battery
terminal(s)/post(s), otherwise serious damage will occur to the electrical system.
Install protective caps on POSITIVE (+) and NEGATIVE (-) battery terminals.