TIGER Instrument User Manual V3.10
Unrivalled Gas Detection. Page 41 of 72 ionscience.com
To add new gases to the table (or delete them), click the Menu button and select Allow
Insert/Delete from the drop-down menu:
An additional row is added at the bottom of the Gas Table into which details of the new gas can be
entered.
Upper and lower alarm levels can be adjusted in the Gas Table. Type the required alarm level in the
appropriate column for the gas in question.
Always save modified gas tables under a different file name. Do not overwrite the original.
For a gas to be available to the instrument, the checkbox next to its name must be
ticked.
To send the Gas Table to the instrument, select Send to Instrument from the drop-down menu
under the Menu button.