Manual No. 0183B 11/29 IPC Series Service Manual
3) Operation
Operation Display
1. Stating Test
X2 X1
2. Scale No. and Decimal point indication mode
- Press the [ TARE ] key ON → (1
st
digit flashes)
- Example: ASIA 3kg (1g) Single range X1=5, X2=4
2.1 Scale No. Setting
- Press the [ * ] key ON four times (
X1=5)
X2 X1
2.2 Decimal point indication Setting
- Press the [ ZERO ] key ON → (2
nd
digit flashes)
X2 X1
- Press the [ * ] key ON three times (X2=4)
X2 X1
- When the C3 mode is required, press the TARE key.
- When finishing, press the Memory Switch to
record the data to the memory, and enter the Ending
Test Mode.
2.1.8 C3 Mode – Span Adjustment
1) Operation
Operation Display
1. Stating Test
2. Span Adjustment mode
2.1 Press the TARE key ON for two times.
2.2 Press the [ * ] key ON → displays original A/D data
(The normal range for original A/D data is 1000 to
25000 counts).
2.3 Press the [ * ] key ON with zero load on the weigh
platter → Approx. 5000 count is diplayed.