Technical Specification
Parameter IT-M3912C-80-240
Current ≤0.05% + 0.05%FS
Load Regulation
Voltage
≤0.01% + 0.01%FS
Current ≤0.05% + 0.05%FS
Output protection
OCP -250A or 250A
OVP 82V
Overpower
protection
-12240W or 12240W
Remote Sense
Compensation
Voltage
≤2V
External analog
optional
current program
External programming voltage -10V to 10V corre-
sponds to current -240A to 240A
current monitor
Current -240A to 240A corresponds to external
monitoring voltage -10V to 10V
voltage program
External programming voltage 0 to 80V Corre-
sponding voltage 0 to 10V
voltage monitor
Voltage 0 to 80V corresponds to external monitor-
ing voltage 0 to 10V
Other parameters:
AC Input
Voltage
3phase 110V520V
Single phase 85V300V
Frequency
50/60Hz
Max. apparent
power
13kVA
Max. input current
25Aac
Max. efficiency
92%
Power Factor 0.99
DC component
≤0.2A
Current harmonics ≤3%
Standard Interface
Standard: USB/LAN/CAN/IO Optional: GPIB/Analog/RS232
Command Re-
sponse Time
0.1ms
Parallel Number ≤16
Working
Temperature
040
Copyright © Itech Electronic Co., Ltd.