Technical Specification
Parameter IT-M3906C-300-60
Read Back
Accuracy
Voltage
≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Voltage ripple
Vp-p ≤300mVpp
Vrms ≤100mV
Setup Temperature
Coefficient
Voltage ≤50ppm/℃
Current
≤50ppm/℃
Read Back Temper-
ature Coefficient
Voltage ≤50ppm/℃
Current
≤50ppm/℃
Rise Time( no load ) Voltage
≤15ms
Rise Time( full load ) Voltage
≤30ms
Fall Time( no load ) Voltage
≤30ms
Fall Time( full load ) Voltage
≤15ms
Dynamic response
time
change from 20% to
90% of rated current
Voltage
≤1ms
Line Regulation
Voltage
≤0.01% + 0.01%FS
Current ≤0.05% + 0.05%FS
Load Regulation
Voltage
≤0.01% + 0.01%FS
Current ≤0.05% + 0.05%FS
Output protection
OCP -63A or 63A
OVP 303V
Overpower
protection
-6120W or 6120W
Remote Sense
Compensation
Voltage
≤3V
External analog
(optional)
current program
External programming voltage -10V to 10V corre-
sponds to current -60A to 60A
current monitor
Current -60A to 60A corresponds to external moni-
toring voltage -10V to 10V
voltage program
External programming voltage 0 to 300V Corre-
sponding voltage 0 to 10V
voltage monitor
Voltage 0 to 300V corresponds to external moni-
toring voltage 0 to 10V
Copyright © Itech Electronic Co., Ltd.