IEEE488.2 Commands
Copyright © Itech Electronic Co., Ltd. 12
Chapter2 IEEE488.2 Commands
*CLS
This order can clean the register as follows:
Standard event status register
Quest condition register
Operation event register
Status byte register
Error code
Order syntax
*CLS
Parameter
None
*ESE
This order can set the parameter of standard event enable register. Setting
parameter can determine which bit value of standard event register is 1 and the
byte will enable ESB of status byte register is 1.
Order syntax
*ESE <NR1>
Parameter
0~255
Reset value
Consult *PSC order
Example
*ESE 128
Quest syntax
*ESE
Return parameter
<NR1>
Reference order
*ESR *PSC *STB