Status commands
Copyright © Itech Electronic Co., Ltd. 19
will beep once.
2. If more than 20 errors occurred, the last one stored in the queue will be
replaced by “-350”, Meaning "too many mistakes.". If the error messages
in the error message queue are not read out, the error following will not
be saved to the queue. If there is no error message, when read error
information, it will return “+0”, meaning “no error”.
3. If turn off the power supply or send “CLS(clear status) order, the error
messages in the error queue will be cleared. *RST order will not clear the error
messages in the error queue.
Order syntax
SYSTem:ERRor?
Parameter
None
SYSTem:REMote
This order is used to set the power supply to remote control mode through the
RS232 interface. Except for the Local key on the front panel, other keys are
locked and can’t be used. Send or receive orders without first sending the order
to configure the remote control may cause unpredictable results.
Order syntax:
SYST:REM
Parameter:
None
Quest syntax:
None
SYSTem:LOCal
This order is used to set the power supply to panel control mode through the
RS232 interface. After execute this order, all the buttons on the front panel can
be used.
Order syntax
SYST:LOC
Parameter
None
Quest syntax
None
SYSTem:RWLock
This order is used to set the power supply to remote control mode through the