EasyManua.ls Logo

ITech IT6723H - Volt:protection:state {0|1|OFF|ON; Volt:protection:triped; Volt:protection:clear

ITech IT6723H
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Commands
Copyright © Itech Electronic Co., Ltd. 28
Unit
V
Query syntax
VOLT:PROT? { MINimum|MAXimum}
Return parameters
<NR2>
VOLT:PROTection:STATe {0|1|OFF|ON}
This order is used to open or close OVP function i.e to set the state of OVP.
Order syntax
[SOURce:]VOLTage:PROTection:STATe
Parameters
0|1|OFF|ON
Query syntax
VOLTage:PROTection:STATe?
Return parameters
0|1
VOLT:PROTection:TRIPed?
This order is used to inquire the execute state of OVP. If the return value is
“1”,this represents that the OVP circuit has been triggered and the OVP state
does not be cleared. If the return value is “0”,then it represents the OVP circuit
does not be triggered.
Order syntax
[SOURce:]VOLTage:PROTection:TRIPed?
Return parameters
0|1
VOLT:PROTection:CLEar
This order is used to clear the state of OVP. After you executing this order, the
output value will be restored to the previous output and the upper limit of OVP
will be kept. Of course, before you sending this order, you should enable the
output voltage lower than the upper limit of OVP. In addition, please note that
making sure to remove the external power supply before you sending this
order.
Order syntax
[SOURce:]VOLTage:PROTection:CLEar
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Related product manuals