Status commands
Copyright © Itech Electronic Co., Ltd. 20
RS232 interface, and LOCAL is not available too. Executing this order will set
the power supply to remote control mode ,the same result as order SYST:REM
function, the only difference is this order will lock all the buttons on the front
panel including LOCAL button.
Order syntax:
SYST:RWL
Parameter:
None
SYSTem:BEEPer
This order is used to test the beeper, after execute this order, the power supply
will issue a beep sound.
Order syntax:
SYSTem:BEEPer[:IMMediate]
Example:
SYST:BEEP
SYSTem:COMMunicate:GPIB:RDEVice:ADDRess
This order is used to set the GPIB communication address.
Order Syntax:
SYSTem:COMMunicate:GPIB:RDEVice:ADDRess <NR1>
Parameter:
0-30
Quest syntax:
SYSTem:COMMunicate:GPIB:RDEVice:ADDRess?
Returned parameter:
<NR1>
SYSTem:INTerface
This order is used to switch the communication interface.
Order syntax:
SYSTem:INTerface <GPIB|USB|RS232>
Parameter:
GPIB|USB|RS232