Load Commands 
  Copyright © Itech Electronic Co., Ltd.  31 
Unit: 
S 
Reset value: 
MINimum 
Query Syntax: 
[SOURce:]INPut:TIMer:DELay? [MINimum|MAXimum] 
Returned value: 
<NR3> 
MEASure:TYPE 
This command is used to set the measurement type of electronic load. 
Command syntax: 
[SOURce:][INPut:]MEASure:TYPE 
Parameter: 
METer|HARMonic|SCOPe 
Reset value: 
METer 
Query Syntax: 
[SOURce:][INPut:]MEASure:TYPE? 
Returned value: 
METer|HARMonic|SCOPe 
RATE:DCCV 
This command is used to set the load rate in CV mode under DC working 
mode.   
Command syntax: 
[SOURce:]RATE:DCCV <SLOW|HIGH> 
Parameter: 
SLOW|HIGH