Measurement Commands 
  Copyright © Itech Electronic Co., Ltd.  9 
Chapter2 Measurement Commands 
MEASure? 
FETCh? 
This command is used to read the measurement value of electronic load. 
The sequence of measurement data is listed in the following table: 
Measurement Value  Description 
CURR_DC  Direct current value 
CURR_RMS  Current effective value 
CURR_MAX  Current maximum value 
CURR_MAXP  The positive peak current 
CURR_MINP  The negative peak current 
VOLT_DC  Direct voltage value 
VOLT_RMS  Voltage effective value 
VOLT_MAX  Maximum voltage 
POW_ACT  Active power 
POW_APP  Apparent power 
POW_REAC  Reactive power 
POW_MAX  Maximum power 
RES  Resistance 
FREQ  Frequency 
CURR_CFAC  Crest Factor 
POW_PFAC  Peak Factor 
VOLT_THD  Total harmonic distortion of voltage 
ETIME  Elapsed time under timing mode 
TEMP  Temperature 
Returned value: 
<NR3> 
MEASure:CURRent? 
FETCh:CURRent? 
This command is used to read the current average value in load input terminal. 
Command syntax:   
MEASure[:SCALar]:CURRent[:DC]? 
FETCh[:SCALar]:CURRent[:DC]?