Performance Verification
1-7
Figure 1-3
Connections for AC volts verification (2MHz frequency only)
NEXT
DISPLAY
PREV
POWER
DCV ACV DCI ACI Ω2 Ω4
FREQ TEMP
REL TRIG STORE RECALL
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
RANGE
AUTO
FILTER MATH
RANGE
2001 MULTIMETER
SENSE
Ω 4 WIRE
HI
INPUT
LO
INPUTS
CAL
500V
PEAK
F
R
FRONT/REAR
2A 250V
AMPS
350V
PEAK
1100V
PEAK
1. 900000 VAC RMS
BNC to dual
banana
50Ω
terminator
Model 2001
5700A Calibrator (Output AC Voltage)
5725 Amplifier (Connect to calibrator)
50Ω Coax
Wideband
output
Ground link installed.
Note: Use internal Guard (EX GRD LED is off).