Performance Verification
1-8
Table 1-3
Limits for normal mode AC voltage verification
Allowable readings (1 year, 18° to 28°C)
2001
ACV
range
Applied
voltage 20Hz 50Hz 1kHz 5kHz 25kHz 50kHz 100kHz 200kHz 1MHz 2MHz**
200mV 190mV 188.716mV
to
191.284mV
189.685mV
to
190.315mV
189.875mV
to
190.125mV
189.875mV
to
190.125mV
189.875mV
to
190.125mV
189.856mV
to
190.144mV
189.647mV
to
190.353mV
189.000mV
to
191.000mV
186.000mV
to
194.000mV
180.100mV
to
199.900mV
2V 1.9V 1.88716V
to
1.91284V
1.89685V
to
1.90315V
1.89875V
to
1.90125V
1.89875V
to
1.90125V
1.89875V
to
1.90125V
1.89856V
to
1.90144V
1.89647V
to
1.90353V
1.89000V
to
1.91000V
1.86000V
to
1.94000V
1.80100V
to
1.99900V
20V 19V 18.8716V
to
19.1284V
18.9685V
to
19.0315V
18.9856V
to
19.0144V
18.9809V
to
19.0192V
18.9742V
to
19.0258V
18.9723V
to
19.0277V
18.9647V
to
19.0353V
18.9000V
to
19.1000V
18.2000V
to
19.8000V
*
200V 190V 188.709V
to
191.291V
189.678V
to
190.322V
189.849V
to
190.151V
189.802V
to
190.198V
189.735V
to
190.265V
189.716V
to
190.284V
189.640V
to
190.360V
***
750V 750V
⎯ 748.12V
to
751.88V
748.72V
to
751.28V
748.49V
to
751.51V
748.12V
to
751.88V
*****
*CAUTION: Do not exceed 2 × 10
7
V•Hz input.
**Use wideband option and connections when performing 2MHz tests.
NOTE: Limits shown do not include calibrator uncertainty. Reading limits do include the adder for AC Coupling of the input.