[SENSe:]CURRent[:DC]:NPLCycles <n>|MIN|MAX
Use this command to select the integration time for DC current measurements. Integration time for DCI
measurements is expressed as the number of power line cycles (NPLC).
Parameters
<n> description
Select the minimum value.
Select the maximum value.
Query
command description
[SENSe:]CURRent[:DC]:NPLCycles?
[SENSe:]CURRent[:DC]:NPLCycles? MIN
[SENSe:]CURRent[:DC]:NPLCycles? MAX
Details
Refer to Integration time and resolution for DC and resistance measurements (on page 3-41) for more
information about NPLC.
[SENSe:]CURRent[:DC]:RANGe <n>|MIN|MAX
Use this command to select the function (DCI) and set a measure range.
Parameters
<n> description
-10 to 10 Select the measurement range in amps.
Select the minimum value.
Select the maximum value.
Query
command description
[SENSe:]CURRent[:DC]:RANGe?
Query the measurement range.
[SENSe:]CURRent[:DC]:RANGe? MIN
[SENSe:]CURRent[:DC]:RANGe? MAX