Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-60 2110-901-01 Rev. C/August 2013
[SENSe:]UNIT <name>
Use this command to select units for temperature measurement.
Parameters
<name> description
Cel
Celsius
Far
Fahrenheit
K
Kelvin
Query
command description
[SENSe:]UNIT?
Query unit sensor type. Returns type.
Details
Select units for temperature measurement. Refer to the following topics in the Temperature measurement (on
page 3-17) section for more information about units: RTD coefficients and temperature calculation (on page 3-
19) and Entering sensor coefficients (on page 3-20)
[SENSe:]VOLTage:AC:RANGe <n>|MIN|MAX
Use this command to select the measurement range for the ACV function.
Parameters
<n> description
0 to 750
Select the measurement range in volts.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]VOLTage:AC:RANGe?
Query the measurement range.
[SENSe:]VOLTage:AC:RANGe? MIN
Query the minimum range.
[SENSe:]VOLTage:AC:RANGe? MAX
Query the maximum range.

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals