3: Basic DMM operation Model 2110 5½ Digit Multimeter
3-20 2110-901-01 Rev. C/August 2013
Entering sensor coefficients
There are three RTD sensor types that allow you to enter coefficient values for sensor parameters.
The coefficients are provided with the documentation supplied by the manufacturer of the RTD.
The following sensors allow you to enter coefficient values:
• USER sensor
• NTCT sensor
• SPRTD sensor
Perform the following steps to enter coefficient values:
1. Press the TEMP key.
2. Press CONFIG, and with SENSOR displayed, press ENTER.
3. Use the â—„ or â–º key to display the sensor type (NTCT, SPRTD, or USER) and press ENTER.
4. Use the â—„ and â–ºkeys to display an RTD parameter and press ENTER.
5. Use the â—„, â–º, â–², and â–¼ keys to enter the coefficient value and press ENTER.
6. Repeat steps 4 and 5 to set the coefficients for all the RTD parameters.
Remote operation
Commands that use the <n> parameter can use the MINimum and MAXimum name parameters. MIN
selects the minimum value and MAX selects the maximum value.
Use the following commands to select the RTD sensor type and set the measurement units:
TEMPerature:RTD:TYPE <name> Select RTD sensor type
<name> = PT100, D100, F100, PT385, PT3916, USER, SPRTD, or NTCT
UNIT <name> Set temperature measurement units
<name> = Cel Celsius
= Far Fahrenheit
= K Kelvin