Model 2110
5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-25
Command Description
:AUTO?
Query auto zero. Returns 0 (off) or 1 (on). ONCE returns 0 (off).
:AVERage
Path to configure average.
:TCONtrol <name>
Specify MOVing or REPeat.
:TCONtrol?
Query average type.
:COUNt <n>|MIN|MAX
Specify count (value).
:COUNt? [MIN|MAX]
Query count.
:STATe <b>
Specify ON or OFF.
:STATe?
Query state. Returns 0 (off) or 1 (on).
STATus subsystem summary
See the following table for the STATus command summary. Refer to the following topics for additional
information on status commands.
Common commands (on page 5-4)
Status model (on page B-1)
Command Description
STATus:
:QUEStionable
Path to configure the Questionable Data Enable Register.
:ENABle <NRf>
Enable bits in the Questionable Data Enable Register. The selected bits
are then reported to the Status Byte.
A 0 value clears the Questionable Data Enable Register.
:ENABle?
Query the Questionable Data Enable Register. The instrument returns a
binary-weighted decimal representing the bits set in the Questionable
Data Enable Register.
:EVENt?
Query the Questionable Data Event Register. The instrument returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
Clear the Questionable Data Event Register.
:PRESet
Clear all bits in the Questionable Data Enable Register.
SYSTem subsystem summary
Command Description
:SYSTem
Path to system commands.
:BEEPer
Issue a single beeping tone.
:STATe <b>
Turn the beeper on or off. Where: 0 = off and 1 = on.
:STATe?
Query state of beeper. Returns 0 (off) 1 (on).
:ERRor?
Query the instrument’s error queue. Up to 20 errors can be stored in the queue.
Errors are retrieved in first-in-first-
out (FIFO) order. Each error string may contain
up to 80 characters.
:LOCal
Place the instrument in the local mode. All keys on the front panel are fully
functional.
:REMote
Place the instrument in the remote mode. All keys on the front panel, except the
LOCAL
key, are disabled.
:VERSion?
Query the present SCPI version.
:DATA:POINts?
Query the number of readings stored in the instrument’s internal memory.

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals