EasyManuals Logo
Home>Keithley>Multimeter>2110-100

Keithley 2110-100 User Manual

Keithley 2110-100
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-24 2110-901-01 Rev. C/August 2013
Command Description
:RSELect?
Query the reference junction type.
:SIMulated <n>|MIN|MAX
Specify the default temperature of the simulated reference
junction.
:SIMulated? [MIN|MAX]
Query the default temperature of the simulated reference
junction.
:REAL <n>|MIN|MAX
Path to offset voltage of the real reference junction.
:REAL? [MIN|MAX]
Query the temperature of the internal reference junction.
:TEMPerature
Path to temperature measurements.
:RTD
Path to resistance temperature detector (RTD).
:TYPE <name>
Specify temperature type: PT100, D100, F100, P385, PT3916,
USER, SPRTD, or NTCT.
:TYPE?
Query the RTD temperature type.
:RZERo <n>|MIN|MAX
Specify the R-Zero coefficient for the user-defined RTD type
(value).
:RZERo?
Query the R-Zero coefficient for the user-defined RTD type.
:ALPHa <n>|MIN|MAX
Specify the alpha coefficient for the user-defined RTD type
(value).
:ALPHa?
Query the alpha coefficient for the user-defined RTD type.
:BETA <n>|MIN|MAX
Specify the beta coefficient for the user-defined RTD type.
:BETA?
Query the beta coefficient for the user-defined RTD type.
:DELTa <n>|MIN|MAX
Set the delta coefficient for the user-defined RTD type (value).
:DELTa?
Query the delta coefficient for the user-defined RTD type.
:SPRTD
Path to the sensor R value at 0 degree Celsius.
:RZERo <n>|MIN|MAX
Specify the sensor R value at 0 degree Celsius (value).
:RZERo?
Query the sensor R value at 0 degree Celsius.
:A4 <n>|MIN|MAX
Specify the A4 coefficient (value).
:A4? [MIN|MAX]
Query the A4 coefficient.
:B4 <n>|MIN|MAX
Specify the B4 coefficient (value).
:B4? [MIN|MAX]
Query the B4 coefficient.
:AX <n>|MIN|MAX
Specify the AX coefficient (value).
:AX? [MIN|MAX]
Query the AX coefficient.
:BX <n>|MIN|MAX
Specify the BX coefficient (value).
:BX? [MIN|MAX]
Query the BX coefficient.
:CX <n>|MIN|MAX
Specify the CX coefficient (value).
:CX? [MIN|MAX]
Query the CX coefficient.
:DX <n>|MIN|MAX
Specify the DX coefficient (value).
:DX? [MIN|MAX]
Query the DX coefficient.
:TRANsducer <name>
Configure transducer. Where: FRTD = 4-wire as the RTD
measurement and RTD = 2-wire as the RTD measurement.
:NTCT
Path to the settings for sensor type NTCT.
:A <n>|MIN|MAX
Specify the A value.
:A? [MIN|MAX]
Query the A value.
:B <n>|MIN|MAX
Specify the B value.
:B? [MIN|MAX]
Query the B value.
:C <n>|MIN|MAX
Specify the C value.
:C? [MIN|MAX]
Query the C value.
:DETector
Path to configure detector.
:BANDwidth <n>|MIN|MAX
Specify bandwidth range (3, 20, 200).
:BANDwidth? [MIN|MAX]
Query bandwidth range.
:ZERO
Path to configure auto zero.
:AUTO <name>
Specify auto zero configuration: OFF, ONCE, or ON.

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals