Model 2110
5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-23
Command Description
:NPLCycles <n>|MIN|MAX
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.06, 1, 2, 10, 100,
MIN, or MAX. Refer to
Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
:NPLCycles?
Query NPLCycles.
:FRESistance
Path to configure fresistance (4-wire
Ω
).
:RANGe <n>|MIN|MAX
Specify range: 0 to 100e6, MIN, or MAX.
? [MIN|MAX]
Query the fresistance range.
:AUTO <b>
Enable or disable auto range. Where, 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
:RESolution? [MIN|MAX]
Query the fresistance resolution.
:NPLCycles <n> MIN|MAX
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.1, 1, 2, 10, 100, MIN,
or MAX.
:NPLCycles? [MIN|MAX]
Query NPLCycles.
:FREQuency
Path to configure frequency.
:VOLTage
Path to configure voltage.
:RANGe <n>|MIN|MAX
Specify range: 0 to 750, MIN, or MAX.
? [MIN|MAX]
Query range.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range setting.
:CURRent
Path to configure current.
:RANGe <n>|MIN|MAX
Specify range.
? [MIN|MAX
Query range.
:APERture <n>|MIN|MAX
Specify the aperture time for the selected function: 0.01, 0.1, 1,
MIN, or MAX.
:APERture? [MIN|MAX]
Query the aperture time for frequency function.
:PERiod
Path to configure period.
:VOLTage
Path to period voltage.
:RANGe <n>|MIN|MAX
Specify range: 0 to 750, MIN, or MAX.
? [MIN|MAX]
Query the period range setting.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range setting.
:CURRent
Path to configure current.
:RANGe <n>|MIN|MAX
Specify range.
? [MIN|MAX
Query the range.
:APERture <n>|MIN|MAX
Specify the aperture time for the selected function: 0.01, 0.1, 1,
MIN, or MAX. Default = 100 ms.
:APERture? [MIN|MAX]
Query the aperture time for period function.
:UNIT <name>
Specify units for temperature measurement (C, F, or K)
:UNIT?
Query units for temperature measurement.
:TCOuple
Path to configure thermocouple.
:TYPE <name>
Specify thermocouple sensor type (B, C, E, J, K, N, R, S, or T).
?
Query thermocouple sensor type.
:RJUNction
Path to thermocouple reference junction.
:RSELect <name>
Specify reference junction type, real or simulated. (REAL or
SIMulated).

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals