Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-62 2110-901-01 Rev. C/August 2013
[SENSe:]VOLTage[:DC]:RANGe <n>MIN|MAX
Use this command to select the measurement range for the DCV function.
Parameters
<n> description
-1000 to 1000
Select the measurement range in volts.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]VOLTage[:DC]:RANGe?
Query the measurement range.
[SENSe:]VOLTage[:DC]:RANGe? MIN
Query the minimum range.
[SENSe:]VOLTage[:DC]:RANGe? MAX
Query the maximum range.
[SENSe:]VOLTage[:DC]:RANGe:AUTO <b>
This command turns autorange on or off for DC voltage measurements.
Parameters
<b> description
0 or OFF
Turn autorange off.
1 or ON
Turn autorange on.
Query
command description
[SENSe:]VOLTage[:DC]:RANGe:AUTO?
Query DC voltage autorange state.