Model 2750 Multimeter/Switch System Service Manual Troubleshooting 4-23
Troubleshooting
Tab le 4- 17
2 signal multiplexing and gain
Range
Signal
(U163)
U129
Pin 1
U129
Pin 8
U129
Pin 9
Gain
(U166)
10
100
1k
10k
100k
1M
10M
100M
S4
S4
S4
S4
S4
S4
S4
S4
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
OFF
OFF
OFF
100
100
10
10
10
1
1
1
Tab le 4- 18
4 signal multiplexing and gain
Range
Signal
(U163)
U129
Pin 1
U129
Pin 8
U129
Pin 9
Gain
(U166)
1
10
100
1k
10k
100k
1M
10M
100M
S4 then S7
S4 then S7
S4 then S7
S4 then S7
S4 then S7
S4 then S7
S4 then S7
S4 then S7
S4 then S7
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
OFF
OFF
OFF
100
100
100
10
10
10
1
1
1
Tab le 4- 19
4 dry circuit signal multiplexing and gain
Range
Signal
(U163)
U129
Pin 1
U129
Pin 8
U129
Pin 9
Gain
(U166)
1
10
100
1k
S4 then S7
S4 then S7
S4 then S7
S4 then S7
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
100
100
100
100

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals