B-26 Calibration Reference Model 2750 Multimeter/Switch System Service Manual
Using the *OPC command
The *OPC (operation complete) command can also be used to detect the completion of
each calibration step. To use *OPC to detect the end of each calibration step, you must do
the following:
1. Enable operation complete by sending *ESE 1. This command sets the OPC (oper-
ation complete bit) in the standard event enable register, allowing operation com-
plete status from the standard event status register to set the ESB (event summary
bit) in the status byte when operation complete is detected.
2. Send the *OPC command immediately following each calibration command. For
example:
:CAL:PROT:DC:STEP1;*OPC
3. Note that you must include the semicolon (;) to separate the two commands, and
that the *OPC command must appear on the same line as the calibration command.
4. After sending a calibration command, repeatedly test the ESB (Event Summary)
bit (bit 5) in the status byte until it is set. (Use either the *STB? query or serial poll-
ing to request the status byte.)
5. Once operation complete has been detected, clear OPC status using one of two
methods: (1) Use the *ESR? query, then read the response to clear the standard
event status register, or (2) send the *CLS command to clear the status registers.
Note that sending *CLS will also clear the error queue and operation complete
status.
Generating an SRQ on calibration complete
An IEEE-488 bus SRQ (service request) can be used to detect operation complete instead
of repeatedly polling the Model 2750. To use this method, send both *ESE 1 and *SRE 32
to the instrument, then include the *OPC command at the end of each calibration com-
mand line, as covered above. Refer to your controllers documentation for information on
detecting and servicing SRQs.

Table of Contents

Other manuals for Keithley 2750

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Summary

Performance Verification

Model 2750 Verification Procedures

Covers verification for DCV, ACV, DCI, ACI, Resistance, Dry Circuit Resistance, Temp, Frequency.

Model 7700 Verification Procedures

Covers verification for Model 7700 plug-in module functions.

Calibration

Calibration Procedures

Details comprehensive, front panel, and remote calibration steps.

Manufacturing and Model 7700 Calibration

Covers specific procedures for manufacturing and Model 7700 calibration.

Routine Maintenance

Troubleshooting

Specific Troubleshooting Steps

Covers checks for display, power supply, digital circuits, signal switching, and errors.

Disassembly

Disassembly and Reassembly Procedures

Step-by-step instructions for disassembling and reassembling the instrument.

Accuracy and Optimization

Accuracy Calculations

Explains how to calculate DC, AC, dBm, and dB accuracy.

Calibration Reference

Calibration Program

Related product manuals