5-2 Disassembly Model 2750 Multimeter/Switch System Service Manual
Introduction
This section explains how to handle, clean, and disassemble the Model 2750. Disassembly
drawings are located at the end of this section.
Handling and cleaning
To avoid contaminating PC board traces with body oil or other foreign matter, avoid
touching the PC board traces while you are repairing the instrument. Some circuit board
areas, especially those under the motherboard shield, have high-impedance devices or sen-
sitive circuitry where contamination could cause degraded performance.
Handling PC boards
Observe the following precautions when handling PC boards:
Wear cotton gloves.
Only handle PC boards by the edges and shields.
Do not touch any board traces or components not associated with repair.
Do not touch areas adjacent to electrical contacts.
Use dry nitrogen gas to clean dust off PC boards.
Solder repairs
Observe the following precautions when soldering a circuit board:
Use an OA-based (organic activated) flux, and take care not to spread the flux to
other areas of the circuit board.
Remove the flux from the work area when you have finished the repair by using
pure water with clean, foam-tipped swabs or a clean, soft brush.
Once you have removed the flux, swab only the repair area with methanol, then
blow dry the board with dry nitrogen gas.
After cleaning, allow the board to dry in a 50°C, low-humidity environment for
several hours.

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A