Model 2750 Multimeter/Switch System Service Manual Calibration Reference B-13
Calibration Reference
AC calibration commands
The :AC commands perform comprehensive (user) calibration of the ACV and ACI func-
tions. Table B-3 summarizes these calibration commands.
Tab le B- 3
AC calibration commands
Command Description
:CALibration
:PROTected
:AC
:STEP1 10mV AC at 1kHz calibration step.
:STEP2 100mV AC at 1kHz calibration step.
:STEP3 100mV AC at 50kHz calibration step.
:STEP4 1V AC at 1kHz calibration step.
:STEP5 1V AC at 50kHz calibration step.
:STEP6 10V AC at 1kHz calibration step.
:STEP7 10V AC at 50kHz calibration step.
:STEP8 100V AC at 1kHz calibration step.
:STEP9 100V AC at 50kHz calibration step.
:STEP10 700V AC at 1kHz calibration step.
:STEP11 100mA AC at 1kHz calibration step.
:STEP12 1A AC at 1kHz calibration step.
:STEP13 2A AC at 1kHz calibration step.

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals