B-12 Calibration Reference Model 2750 Multimeter/Switch System Service Manual
:STEP11
(CALibration:PROTected:DC:STEP11)
Purpose To program the 100mA comprehensive calibration step.
Format :cal:prot:dc:step11 <Cal_current>
Parameters <Cal_current> = 90E-3 to 110E-3 [A]
Description :STEP11 programs the 100mA comprehensive calibration step. The
allowable range of the calibration current parameter is from 90E-3 to
110E-3. Use the 100E-3 value whenever possible for best results.
Example :CAL:PROT:DC:STEP11 0.1 Program 100mA step.
:STEP12
(CALibration:PROTected:DC:STEP12)
Purpose To program the 1A comprehensive calibration step.
Format :cal:prot:dc:step12 <Cal_current>
Parameters <Cal_current> = 0.9 to 1.1 [A]
Description :STEP12 programs the 1A comprehensive calibration step. The allow-
able range of the calibration current parameter is from 0.9 to 1.1. Use a
value of 1 whenever possible for best results.
Example :CAL:PROT:DC:STEP12 1 Program 1A step.

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals