Model 2750 Multimeter/Switch System Service Manual Calibration Reference B-11
Calibration Reference
:STEP8
(:CALibration:PROTected:DC:STEP8)
Purpose To program the 100k 4-wire comprehensive calibration step.
Format :cal:prot:dc:step8 <Cal_resistance>
Parameters <Cal_resistance> = 90E3 to 110E3 []
Description :STEP8 programs the 100k 4-wire resistance comprehensive calibra-
tion step. The allowable range of the calibration resistance parameter is
from 90E3 to 110E3, but 100E3 is recommended for best results.
Example :CAL:PROT:DC:STEP8 100E3 Program 100k step.
:STEP9
(CALibration:PROTected:DC:STEP9)
Purpose To program the 1M comprehensive calibration step.
Format :cal:prot:dc:step9 <Cal_resistance>
Parameters <Cal_resistance> = 900E3 to 1.1E6 []
Description :STEP9 programs the 1M comprehensive calibration step. The
allowable range of the calibration resistance parameter is from 900E3 to
1.1E6. Use the 1E6 value whenever possible, or the closest possible
value.
Example :CAL:PROT:DC:STEP9 Program 1M calibration step.
:STEP10
(CALibration:PROTected:DC:STEP10)
Purpose To program the 10mA comprehensive calibration step.
Format :cal:prot:dc:step10 <Cal_current>
Parameters <Cal_current> = 9E-3 to 11E-3 [A]
Description :STEP10 programs the 10mA comprehensive calibration step. The
allowable range of the calibration current parameter is from 9E-3 to
11E-3. Use the 10E-3 value whenever possible for best results.
Example :CAL:PROT:DC:STEP10 10E-3 Program 10mA step.