Model 2750 Multimeter/Switch System Service Manual Calibration Reference B-7
Calibration Reference
:DATE
(:CALibration:PROTected:DATE)
Purpose To send the calibration date to the instrument.
Format :cal:prot:date <year>, <month>, <day>
Parameters <year> = 2000 to 2099
<month> = 1 to 12
<day> = 1 to 31
Query :cal:prot:date?
Response <year>, <month>, <day>
Description The :DATE command allows you to store the calibration date in instru-
ment memory for future reference. You can read back the date from the
instrument over the bus by using the :DATE? query or the CALIBRA-
TION selection in the front panel CAL menu.
NOTE The year, month, and day parameters must be delimited by commas.
Example :CAL:PROT:DATE 2001,3,16 Send cal date (3/16/2001).
:CAL:PROT:DATE? Request cal date.
:NDUE
:CALibration:PROTected:NDUE)
Purpose To send the next calibration due date to the instrument.
Format :cal:prot:ndue <year>, <month>, <day>
Parameters <year> = 2000 to 2099
<month> = 1 to 12
<day> = 1 to 31
Query :cal:prot:ndue?
Response <year>, <month>, <day>
Description The :NDUE command allows you to store the date when calibration is
next due in instrument memory. You can read back the next due date
from the instrument over the bus by using the :NDUE? query or the
front panel CAL menu.
NOTE The next due date parameters must be delimited by commas.
Example :CAL:PROT:NDUE 2002,3,16 Send due date (3/16/2001).
:CAL:PROT:NDUE? Request due date.