EasyManua.ls Logo

Keithley 3730

Keithley 3730
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
(440) 248-0400
www.keithley.com
Specifications are subject to change without notice.
Page 2 of 3 SPEC-3722 Rev. A / September 2007
Model 3722
Dual 1x48, High Density Multiplexer Specifications
96 differential channels, 300V Volts/1 Amp
GENERAL:
ACTUATION TIME: 4ms
RELAY TYPE: Latching electromechanical
RELAY DRIVE SCHEME: Matrix
OPERATING ENVIRONMENT:
Specified for 0ºC to 50ºC.
Specified to 70% R.H. at 35ºC
STORAGE ENVIRONMENT: -25ºC to 65ºC
WEIGHT: 2.5lb
SAFETY: Conforms to European Union Directive 73/23/EEC,EN61010-1
EMC: Conforms to European Union Directive 2004/108/EC,EN61326-1
Typical Scanning Speeds, switch only
3
Ch/Sec
Sequential scanning ,single channel, immediate trigger advance >120
Typical Scanning Speeds, with measurements into memory
4
Ch/Sec
DCV (10V range) or 2W Ohms (1K range) >110
3 or 4 Wire RTD >100
4 Wire Ohms (1k range) >100
ACV (10V, 400Hz range) >110
ACCESSORIES AVAILABLE:
3791-KIT104-R 104 pin male D-sub connector kit (Contains 2 male D-sub connector with housings and 208
solder-cup contacts)
3722-MTC-1.5 104 pin D-sub male to male cable, 1.5m.
3722-MTC-3 104 pin D-sub male to male cable, 3m.
3791-CIT Contact insertion and extraction tool.
3
Scanning script local to mainframe, within same bank, break before make.
4
3706 mainframe with autorange off, limits off, DMM.autodelay=0, dmm.autozero=0, 4 ½ digits ( NPLC=.006), for ACV
dmm.detectorbandwidth=300,for OHMs dmm.offsetcompensation=off, dmm.opendetector=off. Scanning script local to
mainframe, sequential scan within same bank (2 pole) or card (4 pole), and break before make switching.