340 OUTPUT 708 ?F3R4Ul9E-9X ! 263; program for 13nC.
350 OUTPUT 727 YC0X
! 617; disable zero check.
360 WAIT 2000
370 OUTPUT 727 ;"NiX-
! 617; enable suppress.
380
OUTPUT 708
; ~OlX' ! 263;
output 20nC to 617.
390 WAIT
2500
400 BEEF
2,200
0 OUTPUT 727
Yfi19E-9X" ! 617: send cal value.
410 WAIT 2000
420 PRINT 'COUL RANGES CALIBRATED:-
430 PRINT '20nC' 0 PRINT
440 OUTPUT 708 i
'FZRlUlSOE-3X' ! 263; program for 190mV.
450 DATA .19,1.9,19
460 FOR I=1 TO 3
470 OUTPUT 727 ;'CIZONOFOR";I;"X"
! 617; select volts range.
480 OUTPUT 708 ;"R~;I;YZIOIX" ! 263; select range, output 0U.
430 OUTPUT 727 ;YZlXC0X
! 617; enable :. COT. and disable z. chk.
500 OUTPUT 727 YNIX'
! 617; enable suppve5~.
510 OUTPUT 708 ;'Z0X' ! 263; output programmed voltage.
520 READ U
530 WAIT 2000 0 CLEAR
540 BEEP 2,200 0 OUTPUT 727 ;mAu;U;"X~ ! 617; send cal value.
550 WAIT 2000
560 NEXT I
570 OUTPUT 727 ; 'Z0N0R4X"
! 617; select 200V range.
580 OUTPUT 708 ;"FSOIXw ! 263; source ert U to 617.
590 WAIT 2000
600 BEEP 2,200 0 OUTPUT 727 ;"A190X" ! 617; send cal value.
610 WAIT 2000
620
OUTPUT 727 ;"ClFZR8X' 1 617: select 20Gohm range.
630 PRINT 'VOLTS RANGES CALIBRATED:'
640 PRINT "All" 0 PRINT
650 OUTPUT 708 ;"FOREWIX"
! 263; select 10Gohm range, Quard on.
660 BEEP 0 DISP "SET 617 GUARD SWITCH TO 'ON' POSITION."
670 DISP 0 DISP 'PRESS 'CONT' TO CONTINUE."
680 PAUSE
690 FOR I=8 TO 2 STEP -1
700 IF I-8 THEN 770
710 IF I<7 THEN 770
720 CLEAR
730 BEEP @ DISP "SET 617 GUARD SWITCH TO 'OFF' POSITION."
740 DISP 0 DISP "PRESS 'CONT' TO CONTINUE."
750 PAUSE
760 CLEAR @ OUTPUT 708 ;"W0X" ! 263; disable guard.
770 OUTPUT 708 ;"R';I"XZlOlX" ! 263; output
0ohms
to 617.
780 OUTPUT 727 ;"ClZQNQR";I;"X" ! 617; select FanQe.
730 OUTPUT 727 ;"ZlXCQXNlX" ! 617; zero display.
800 OUTPUT 708 ;"ZQX" ! 263; source programmed resistance.
617 CALIBRATION PROGRAM (Cont.)