Series 2600 System SourceMeters Reference Manual DUT Test Connections 3-19
Return to Section 3 topics 2600S-901-01 Rev. A / May 2006
Selecting the Output-off state
Output-off state menu
The OUTPUT configuration menu can be accessed by pressing the CONFIG key
and then the appropriate OUTPUT ON/OFF key. In the configuration menu, select
OFF STATE to display the OUTPUT OFF STATE menu.
With the OUTPUT OFF STATE menu displayed, select the desired output-off
state: HI-Z (high-impedance), NORMAL or ZERO.
Remote programming
Table 3-3 lists the commands to select the Output-off state.
NOTE The OUTPUT OFF STATE menu can also be
accessed by navigating the configuration menu that
is displayed by pressing the CONFIG key.
Table 3-3
Commands to select Output-off state
Command
*
Description
smuX.source.offlimiti = ivalue Sets 2611/2612 current limit in
normal Output-off state.
smuX.source.offmode = smuX.OUTPUT_NORMAL Selects normal Output-off state.
smuX.source.offmode = smuX.OUTPUT_HIGH_Z Selects high-impedance Output-
off state.
smuX.source.offmode = smuX.OUTPUT_ZERO Selects zero Output-off state.
* Model 2601/2611: smuX = smua
Model 2602/2612: smuX = smua (Channel A) or smub (Channel B).

Table of Contents

Other manuals for Keithley SourceMeter 2612

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2612 and is the answer not in the manual?

Keithley SourceMeter 2612 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2612
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals