12-128 Instrument Control Library Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 12 topics
status.system3.*
status.system3.condition
status.system3.enable
status.system3.event
* = condition, enable or event
Attribute TSP-Link system3 data structure register set.
Usage Reads condition, enable and event registers:
enablereg = status.system3.condition
enablereg = status.system3.enable
enablereg = status.system3.event
Writes to enable register:
status.system3.enable = enablereg
Set enablereg to one of the following values:
0 Clears all bits.
1 or status.system3.EXTENSION_BIT Sets EXT bit (B0).
1 or status.system3.EXT Sets EXT bit (B0).
status.system3.NODEn Sets a node bit (Bn); n = 29 to 42.
Remarks
In an expanded system (TSP-Link), this attribute is used to read or write to the
system3 node registers.
Reading a system3 node register returns a numeric value whose binary equivalent
indicates which register bits are set. The bits of the system3 node register are
identified as follows:
Bit B0 – EXT bit Bit B4 – Node 32 Bit B8 – Node 36 Bit B12 – Node 40
Bit B1 – Node 29 Bit B5 – Node 33 Bit B9 – Node 37 Bit B13 – Node 41
Bit B2 – Node 30 Bit B6 – Node 34 Bit B10 – Node 38 Bit B14 – Node 42
Bit B3 – Node 31 Bit B7 – Node 35 Bit B11 – Node 39 Bit B15 – Not used
For example, assume value 9 is returned for the enable register. The binary
equivalent is 0000000000001001. This value indicates that bit 0 (EXT) and bit 3
(Node 31) are set.
Assigning a value to the status.system3.enable attribute sets the extension bit or
a node bit of the system3 node enable register.
Details See “System Summary Event Registers” in Appendix D.
Also see status.system.*, status.system2.*, status.system4.*, status.system5.*
Example Sets the extension bit of the system3 enable register:
status.system3.enable = status.system3.EXT

Table of Contents

Other manuals for Keithley SourceMeter 2612

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2612 and is the answer not in the manual?

Keithley SourceMeter 2612 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2612
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals