Series 2600 System SourceMeters Reference Manual Status Model D-17
Return to Appendix D topics 2600S-901-01 Rev. A / May 2006
Status register sets
As shown in Figure D-1 through Figure D-5, there are five status register sets in
the status structure of the SourceMeter; System Summary Event Status, Standard
Event Status, Operation Event Status, Measurement Event Status, and Question
-
able Event Status.
System Summary Event Registers
As shown in Figure D-1 and Figure D-2, there are five register sets associated
with System Event Status. These registers summarize system status for various
nodes connected to the TSP-Link (see Section 9). Note that all nodes on the TSP-
Link share a copy of the system summary registers once the TSP-Link has been
initialized. This feature allows all nodes to access the status models of other
nodes, including SRQ.
In a TSP-Link system, the status model can be configured such that a status
event in any node in the system can set the RQS (Request for Service) bit of the
Master Node Status Byte. See
TSP-Link system status” in this appendix for
details on using the status model in a TSP-Link system.
Commands for the system summary registers are summarized in Table D-5.
For example, either of the following commands will set the EXT enable bit:
status.system.enable = status.system.EXT
status.system.enable = 1
When reading a register, a numeric value is returned. The binary equivalent of this
value indicates which bits in the register are set. For details, see “
Reading regis-
ters” in this appendix. For example, the following command will read the system
enable register:
print(status.system.enable)
The used bits of the system event registers are described as follows:
Bit B0, Extension Bit (EXT)Set bit indicates that an extension bit from
a another system status register is set.
Bits B1-B14*, NODEn — Indicates a bit on TSP-Link node n has been set
(n =1 to 64).
* status.system5 does not use bits B9 through B15.

Table of Contents

Other manuals for Keithley SourceMeter 2612

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2612 and is the answer not in the manual?

Keithley SourceMeter 2612 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2612
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals